The Parsian Medical OEM Oximetry Modules
- Parsian Medical OEM modules amplify and digitize the signal in the sensor, resulting in a stronger and higher quality signal being received by the host system.
- Offering several sizes that can serve as drop-in replacements for Nellcor OxiMax™ and Masimo SET and Dolphin One OEM modules.
SPECIFICATIONS
Evironmental
Operating temp: 23 F to 113 F (-5 C to +45 C)
Storage temp: -4 F to 140 F (-20 C to + 60 C)
Operating humidity: 5% to 95%, non-condensing
Operating altitude: -1,000 ft to 18,000 ft (-304 m to 5,486 m) (503 mbar to 1,059 mbar)
Range
SpO2 (Functional): 0-100%
Pulse rate (bpm): 30-240
Normal sensitivity (Perfusion %): 0.6%-20%
High sensitivity (Perfusion %): 0.2%-20%
Body temperature: 0-60°C
Resolution
SpO2 (functional): 1%
Pulse rate: 1 bpm
Body temperature: 0.1°F or °C
Modes
Averaging: Smart algorithm utilizes 4 to 12 second data stream depending on signal to noise ratio
Sensitivity: Normal and High
Patient types: Adult / Pediatric / Neonatal
Power
Voltage input: +5V
Nominal power consumption: 0.50 – 0.75W
Connectors
Power / Communication / Reset connector
Sensor connector
Standard Interface for Clinical Use
Streaming % SpO2, Pulse Rate
Streaming Temperature
Streaming photo-plethysmogram (27.5 hz)
Streaming sensor condition, system failure
Sensors
Adult and Ped > 30 kg
Adult and Ped > 30 kg
Ped and Neo ≤ 30 kg
DESCRIPTION
Oximetry performance is the most critical right when conditions are most difficult. Parsian’s pm 2019 OEM modules are designed to substantially improve performance and eliminate nuisance alarms associated with motion and low perfusion.
PM 2019
Size: 3.5 x 2.0 inches (89 x 51 mm)
Weight: 1.0 ounces (28g)
PCB thickness: 0.062 +/- 0.007 in
Top clearance: 0.19 in
Bottom clearance: 0.05 in
Mounting hole diameter: 0.125 in
Drop-in replacement for the Masimo MS-5
ACCURACY
SpO2 (functional) | |||
---|---|---|---|
Adult or Pediatric | No Motion and Normal Perfusion | 70 -100 % | ± 2 % |
0 – 69 % | Unspecified | ||
Neonate | No Motion and Normal Perfusion | 70 -100 % | ± 3.5 % |
0 – 69 % | Unspecified |
Pulse Rate (bpm) | Adult, Pediatric, or Neonate | No Motion and Normal Perfusion | 30 – 240 bpm | ± 3 bpm |
---|---|---|---|---|
SpO2 (functional) | Adult or Pediatric | Motion or Low Perfusion<0.2 % | 70 -100 % | ± 3 % |
0 – 69 % | Unspecified | |||
Pulse Rate (bpm) | Adult or Pediatric | Motion or Low Perfusion<0.2 % | 30 – 240 bpm | ± 5 bpm |
Temperature (°C) | 32.0 – 42.0 | 0.1°C | ||
0.0 – 31.9 42.1 – 60.0 |
0.2°C |
Do You Have Any Question?
If You Have Any Question About OEM Module Please Ask Us.